logo
Chào mừng đến Shaanxi KeGu New Material Technology Co., Ltd
8616602956098

Compact Nanoparticle Size & Zeta Potential Analyzer

Basic Properties
Place of Origin: Trung Quốc
Brand Name: KeGu
Model Number: Máy phân tích kích thước hạt nano & tiềm năng Zeta
Trading Properties
Minimum Order Quantity: Có thể thương lượng
Price: Price is negotiable
Payment Terms: L/C,D/A,D/P,T/T,Western Union
Supply Ability: Có thể thương lượng
Tóm tắt sản phẩm
Nanoparticle Size & Zeta Potential Analyzer Product Overview The NKT‑N9 series nanoparticle size analyzers are specifically designed for the characterization of nanoparticle size and size distribution. Based on the principles of Dynamic Light Scattering (DLS) and Photon Correlation Spectroscopy (PCS...

Chi tiết sản phẩm

Làm nổi bật:

Zeta Potential Analyzer

,

Nanoparticle Size Zeta Potential Analyzer

Mô tả sản phẩm
Nanoparticle Size & Zeta Potential Analyzer
Product Overview

The NKT‑N9 series nanoparticle size analyzers are specifically designed for the characterization of nanoparticle size and size distribution. Based on the principles of Dynamic Light Scattering (DLS) and Photon Correlation Spectroscopy (PCS) technology, the instrument uses a laser to illuminate nanoparticles uniformly dispersed in a sample cell.

The detection fiber receives, at a 90° angle, the dynamic scattered light whose intensity fluctuates over time due to the Brownian motion of the sample particles. This light is transmitted to a photomultiplier tube (PMT), which converts the optical signals into discrete pulse signals. A digital correlator then performs autocorrelation calculations to obtain the autocorrelation curve of the sample. Combined with mathematical methods, the diffusion coefficient of the particles is derived, and finally the particle size and size distribution are calculated according to the Stokes‑Einstein equation.

Scope of Application

The NKT‑N9 laser particle size analyzer is widely used for particle size characterization as well as dispersion stability and system stability analysis of organic or inorganic nano‑emulsions, powders, polymers, micelles, virus antibodies, proteins, and other samples in fields such as:

  • New materials
  • Medicinal products
  • Bioengineering
  • Life sciences
  • Chemical engineering
  • Coatings
  • Food safety
Product Parameters
Parameter Specification
Measurement Principle Dynamic Light Scattering (DLS), Photon Correlation Spectroscopy (PCS), Static Light Scattering (SLS)
Particle Size Range 0.5 nm – 10,000 nm (depending on sample options)
Detection Angle 90°
Test Parameters Particle size, molecular weight, second virial coefficient
Laser Integrated temperature control system, max power 50 mW, wavelength 532 nm or 405 nm solid-state laser (optional)
Correlator 512 physical channels, sampling time dynamically adjustable from 0.1 µs to 1 ms
Detector Photomultiplier Tube (PMT)
Filtering Method Optical trap filtering
Accuracy Error Better than ±1% (average particle size, NIST traceable reference material)
Repeatability Error Better than ±1% (average particle size, NIST traceable reference material)
Sample Cell 12 mm quartz cuvette, 1 ml – 4 ml
Concentration Range 0.1 mg/ml – 40 % w/v (depending on sample options)
Molecular Weight Range 340 Da – 2 * 10⁷ Da (depending on sample options)
Measurement Speed Less than 5 minutes per run
Temperature Control Range 5 °C – 90 °C
Temperature Accuracy ±0.1 °C
Intensity Adjustment Range 0 – 100%, automatic adjustment
Environmental Requirements 5 °C – 45 °C, 10% – 85% relative humidity (non‑condensing)
Power Supply AC100 V – 240 V @ 50 Hz – 60 Hz, DC12 V/10 A
Dimensions 440 mm * 245 mm * 228 mm
Instrument Weight 9.5 kg
Applicable Standards GB/T 29022‑2021 / ISO 22412‑2017, GB/T 19627‑2005 / ISO 13320‑1996
Model Comparison
Feature NKT‑N9 NKT‑N9H
Light Source Fiber‑coupled semiconductor laser Fiber‑coupled semiconductor laser
Wavelength λ = 520 nm Wavelength 1: λ = 520 nm, P = 30 mW
Wavelength 2: λ = 405 nm, P = 20 mW
Power P = 30 mW
Applications Suitable for most nanoparticle tests, but cannot effectively test light‑absorbing samples (e.g., red samples) Suitable for effective testing of all nanoparticles, especially colored samples in industries such as dyes and pigments, enabling effective particle size testing for full‑color‑range samples
Technical Features
All‑Fiber Test Optical Path – High Sensitivity

The test optical path, constructed with a high‑performance, low‑noise fiber‑coupled semiconductor laser and an all‑fiber dynamic scattered light receiving system, uses a low dark count, high‑sensitivity HAMAMASTU photomultiplier tube (PMT) as the photodetector. This gives the entire test system extremely high sensitivity and signal‑to‑noise ratio, laying the foundation for test accuracy and high resolution.

High‑Speed Digital Correlator – High Resolution

The independently developed high‑speed digital correlator is used as the core device for signal processing. It acquires dynamic scattered light intensity in real time at high speed and performs autocorrelation calculations. It features a signal resolution capability of 6 ns, a data processing speed of hundreds of megahertz, and a linear dynamic range greater than 10¹¹, ensuring the integrity of autocorrelation curves for particles of different sizes across the full measurement range.

Full‑Range Calibration and Validation – Ensuring Test Accuracy

Multiple national particle size reference materials are used for calibration and validation across the full measurement range, with test accuracy errors all below 1 %.

High‑Precision Temperature Control

The semiconductor laser, equipped with automatic temperature control technology, ensures stable and reliable laser output. The high‑precision temperature control system keeps the tested sample in the sample cell at a constant temperature throughout the measurement process, avoiding test deviations caused by changes in medium viscosity and particle Brownian motion velocity due to temperature variations, thereby ensuring reliable and stable test results.

One‑Click Dual Wavelength Switching – Wider Applicability

Using dual‑wavelength (λ1 = 520 nm, λ2 = 405 nm) fiber‑coupled semiconductor lasers as the light source overcomes the limitation that a single‑wavelength laser cannot test light‑absorbing samples, enabling effective particle size testing for full‑color‑range samples. One‑click switching between the two wavelengths makes the instrument applicable to a wider range of samples.

For further information or to discuss your specific requirements, please contact us directly.
Sản phẩm liên quan

Gửi Yêu Cầu